| Separate Electrodes |
|
|
SM21-AG6 Heavy Duty Measuring |
SM21-AL6 Heavy Duty/ High Temp. |
 |
 |
SC21-AGP26 Combo pH/Ref |
|
SR20-AP26 Double Junction Reference |
SR20-AC22 High Temp. Reference |
SM60-T1 Pt1000 RTD |
|
| |
- SM21’s are pH measuring or glass electrodes.
G is general purpose glass while L denotes high temperature (130 C).
4 is bulb shape while 6 is Heavy Duty dome shaped glass. |
| -SR20’s are reference electrodes. |
| SR20-AP26 has a double junction to slow down poisoning of the element. |
|
|
| SR20-AC22 has pelletized KCl to perform up to 122C |
|
|
-SC21’s are combination pH/Reference
The SC21-AGP26 has Heavy Duty glass and double junction reference. |
|
|
-SM60’s are temperature RTD’s.
SM60-T1 is Pt1000 which is recommended.
SM60-T2 for customers that have to have Pt100 for one reason or another. |
|
|